1.Anaesthesia system
Flow meter
Adjusting range
O2 0.1~1 L/min、1~10 L/min
N2O 0.1~1 L/min、1~10 L/min
Accuracy (20℃ 101.3kPa)
Above 1L/min ±10% of indicated value
Below 1L/min Class 4
Interlocked control of O2 and N2O flows O2 and N2O flow meters interlocked for
proportional adjustment. O2 concentration be ≥25 %.
Quick oxygen supply
Quick oxygen supply flow 35 L/min~ 75 L/min
Enflurane evaporator and isoflurane evaporator
Concentration adjusting range 0.5%~5%
With temperature and flow self compensation functions
Oxygen emergency alarm and laughing gas cut-off protection
When O2 pressure is lower than 0.2 MPa, an alarm will be given lasting at least for 7s and N2O supply will be cut off.
2.Anaesthesia ventilation system:
Ventilation mode:
——IPPV
——SIMV
——SIPPV
——SIGH
——MANU
Preset range of anaesthesia ventilation parameter:
— Frequency (Except SIMV) 6 ~ 60 /min
—Frequency(SIMV) 1 ~ 12 /min
— Inhalation and exhalation ratio (TI:TE) (Except SIMV) 2:1 1:1 1:1.5 1:2 1:2.5 1:3
—Inhaling time(SIMV) 0.2 ~ 6 s
—Tidal volume (VT) 0.05 ~ 1.50 L
—TIP:TI 0 ~ 50%
— Inhalation triggering pressure -0.2 ~﹢2 kPa
Monitoring range of ventilation parameter
Fig display:
— Tidal volume (VT) 0.03 ~ 1.50 L
— Frequency (Freq) 1 ~ 99 /min
— Minute ventilation (MV) 0 ~ 99L/min
— Airway peak pressure (Pmax) 0 ~ 9.9 kPa
—Inhaling time (TI) 0 ~ 9.9s
— Inhalation oxygen concentration 0 ~ 100%.